Loading...

Vaporizer

Vapo-ALPHA

by J&TEC posted Apr 01, 2016
vapo-alpha.jpg

 


VAPO-ALPHA

Temperature compensated
International color-coding
Compatible with any machine
Interlocking system with selectatec mount

 


vapo-title.jpg

 

Technical specification

Dimension

Selectatec : 126mm (W) × 192mm (D) × 230mm (H)

Cage : 135mm (W) × 200mm (D) × 230mm (H)

 

Operation Environment

Temperature : 15 ~ 35 ℃

Humidity : 30 ~ 75 %

 

Agent capacity

Charged : 220mL of free volatile anesthetic agent

Charging : 300mL(nominal) to charge by vaporizers

 

Calibration

The dials of the Vapor-ALPHA vaporizers for enflurane, isoflurane, and halothane are calibrated to 6% v/v in steps of 0.2% up to 1%, and in steps of 0.5% from 1% to 5% at 21℃ using O₂at 5L/min as the carrier gas

top_filler.jpg

 

 

Calibration of Sevoflurane

The dial of the Vapor-ALPHA vaporizer for sevoflurane is calibrated to 7% v/v

Dial is marked in steps of 0.2% up to 1% v/v, and in steps of 1% between 1% v/v and 7%

 

Flow rate

Vaporizer are designed to provide consistent output throghout clinical flow range from 200mL/min to 15L/min

 

Vaporizer type

Agent : Halothan, enflurane, Isoflurane, sevofluran

Filler type : Top filler, Quik filler(sevoflurane)

Mounting system : Cagemount, selectatec(with interlock)

 

Operating

Temperature : 18℃ to 35℃

Humidity : 30% to 75% relative humidity

quik_filler.jpg

 